FYRA Vacuum Gauge for End of Primary Drying Detection UL/CSA/CE certified
FYRA for End of Primary Drying Detection. This bundle is set up to allow for vacuum drying applications like freeze-drying and vacuum ovens to help determine the end of drying. Will help you detect when most of the water or residual solvent from a product has been removed or the end of primary drying has been reached in a vacuum drying application. Background on How this FYRA End of Primary Drying package works: When a freeze dryer is equipped with both a thermal and direct measurement, these sensors can be utilized to indicate the end of primary drying. Step One: Clean process documentation | Get a baseline Characterize the convection-enhance pirani and the capacitance manometer (in this case the DCP Quantum) against each other in a dry, empty and refrigerated system. The freeze dryer is turned on, as per the manufacturer’s operator manual, with nothing on the shelf. The shelf temperature can be left at room temperature, and if so the cold trap, the condenser, and the vacuum pum